Model Number 8410-0408
Medical Device Problem Codes Stretched (1601); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/24/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lot number was provided.A review of the approved device history records indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The pusher and implant coil were returned for analysis.The implant was not attached to the coil.The hypotube was noted to be bent at the proximal section.The transition section between the body coil and the hypotube was stretched.The distal portion of the pusher body coil was noted to be within specifications.The implant was stretched at proximal section, and notably damaged at the distal portion.Based on the available information and evaluation of the returned device, the complaint can be confirmed.The root cause cannot be determined; however, the device exhibits evidence that it was subjected to tensile forces that exceeded its strength specifications.The instructions for use (ifu) identifies premature or difficult coil detachment as potential complications associated with use of the device.
 
Event or Problem Description
It was reported that during the third attempt to reposition the embolization coil in the aneurysm, the coil stretched and then detached.The entire coil was removed from the aneurysm by aspiration.There was no reported patient injury.The patient was reported to be doing well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYPERSOFT 3D ADVANCED
Common Device Name
EMBOLIZATION COIL
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise drive
aliso viejo CA 92656
MDR Report Key8011779
Report Number2032493-2018-00223
Device Sequence Number1262310
Product Code HCG
UDI-Device Identifier00816777021304
UDI-Public(01)00816777021304(11)180125(17)221231(10)1801251W5
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K131948
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date12/31/2022
Device Model Number8410-0408
Device Lot Number1801251W5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/27/2018
Initial Report FDA Received Date10/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-