• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. COMPLETE; SOLUTIONS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. COMPLETE; SOLUTIONS Back to Search Results
Model Number 09424X
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Codes Unspecified Infection (1930); Visual Impairment (2138); Visual Disturbances (2140)
Date of Event 06/14/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Alternative report identification number: (b)(4).Device evaluation: the solution was not returned to the manufacturing site for investigation.Therefore, the consumer's reported complaint could not be confirmed.Manufacturing record review: a review of the records was performed.The records for production process were found to be acceptable, all testing items were completed and met specifications.There was no non-conformances related to this complaint.The reported lot was deemed acceptable for release and there was no indication of a malfunction or product quality deficiency.Labeling review: the directions for use (dfu) were reviewed.The dfu adequately provides instructions, precautions, and warnings for the proper use and handling of the product.Based on the results of the investigation, a product deficiency was not determined.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported through a medwatch report that a consumer purchased complete multi purpose solution.The consumer brought it home, broke the seal, poured it into the contact case, and washed out their contacts.The lens case was put on the bathroom shelf.The consumer used the contact solution for several weeks and started developing sores in the right and left eyes.They went to see the eye doctor and was referred to an ophthalmologist.The ophthalmologist cultured the solution and found that it grew pseudomonas aeruginosa, an infection that is methicillin resistant.The consumer has very little to no vision in their left eye, and in the right eye it's hazy, shadowy.The consumer has been restricted by their doctor from driving and working.Consumer has been treated to hopefully restore some of their vision.The doctor stated they are going to have a permanent scar across the pupil and likely will not regain a lot of sight in the left eye.Although that is not yet certain.No further information was provided.Medications prescribed: tobramycin eye drops, vancomycin eye drops, prednisone steroid eye drops.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPLETE
Common Device Name
SOLUTIONS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key8012217
Report Number3004178847-2018-00011
Device Sequence Number1585376
Product Code LPN
UDI-Device Identifier30827444000615
UDI-Public(01)30827444000615(17)190618(10)ZC06600
Combination Product (Y/N)Y
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K014202
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/02/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date06/18/2019
Device Model Number09424X
Device Lot NumberZC06600
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 07/11/2018
Supplement Date Received by Manufacturer10/25/2020
Initial Report FDA Received Date10/28/2018
Supplement Report FDA Received Date11/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Disability;
Date Report Sent to FDA01/01/2005
-
-