Model Number 1884068HS
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis results were not available as of the date of this report.A follow-up report will be submitted when analysis is complete.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A health care provider (hcp) reported via a manufacturer representative that when the surgeon started the shaver prior to the procedure, the bur was broken.A different bur was used for the procedure.There was no patient involvement.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received from the health care provider (hcp) reported via a manufacturer representative that there was no delay in the case, just a change of the bur for dacryocystorhinostomy procedure.
 
Additional Manufacturer Narrative
Analysis found that the inner shaft broke 0.625¿ from the distal face of the inner hub.There were striations around the outside diameter of the break point indicating metal on metal contact during use.The break point corresponds to the proximal end of the outer tube in the front hub.There was no damage to the distal tip.When viewed under magnification, there was deformation and indentations of the front hub locking area.There were no bends or signs of a concentricity issues.There was no indication of device fragments.There were no signs of heat deformation or discoloring.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® BUR - HIGH SPEED CURVED
Common Device Name
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key8013932
Report Number1045254-2018-00542
Device Sequence Number16459477
Product Code EQJ
UDI-Device Identifier20681490692919
UDI-Public20681490692919
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/02/2022
Device Model Number1884068HS
Device Catalogue Number1884068HS
Device Lot Number0208716334
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/02/2018
Supplement Date Received by Manufacturer10/31/2018
12/14/2018
Initial Report FDA Received Date10/29/2018
Supplement Report FDA Received Date11/21/2018
12/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-