• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO U SERIES ANGLED LONG AM" ATTACHMENT"; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO U SERIES ANGLED LONG AM" ATTACHMENT"; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE Back to Search Results
Catalog Number 5100010072
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr report is part of the malfunction summary reporting program, exemption number e2015055.Reported events - 2 events were reported for this quarter.  product return status - 2 devices were received for evaluation.Evaluation status - 2 device evaluations are in progress.  additional information - 2 devices were not labeled for single-use.2 devices were not reprocessed and reused.
 
Event or Problem Description
This report summarizes 2 malfunction events in which the device reportedly overheated.2 events had no patient involvement; no patient impact.
 
Event or Problem Description
This report summarizes 2 malfunction events in which the device reportedly overheated.2 events had no patient involvement; no patient impact.
 
Additional Manufacturer Narrative
This mdr report is part of the malfunction summary reporting program, exemption number (b)(4).Supplemental rationale corrected data.2 total events were initially reported.2 previously reported events are included in this follow-up record.  product return status devices were received.  evaluation status confirmed 2 reported events were confirmed during testing.- 1 device was found to be affected by corrosion.- 1 device was found to be affected by lubrication breakdown.  additional information 2 devices were not labeled for single-use.2 devices were not reprocessed and reused.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
U SERIES ANGLED LONG AM" ATTACHMENT"
Common Device Name
DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
MDR Report Key8014257
Report Number0001811755-2018-01883
Device Sequence Number19174656
Product Code ERL
UDI-Device Identifier04546540074546
UDI-Public04546540074546
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized2
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number5100010072
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/02/2018
Supplement Date Received by Manufacturer06/02/2018
Initial Report FDA Received Date10/29/2018
Supplement Report FDA Received Date01/31/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-