• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY STRATA II VALVE, REGULAR; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY STRATA II VALVE, REGULAR; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Model Number 42866
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/25/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported the patient had cerebrospinal fluid shunt surgery on (b)(6) 2018.Within a half a month, it was found that the cerebrospinal fluid in the patient was clogged.On (b)(6) 2018, the doctor explanted and replaced the valve in the patient's body with a new valve, and the postoperative shunt was normal.The patient's status at the time of the report was alive-no injury.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
The returned valve was patent.It met the requirements for valve flux, siphon, reflux, pressure-flow, and pre-implantation testing.However, the valve did not meet the requirement for leak testing.The valve did not meet the leak test requirements due to a cut in the dome.It is unknown how or when the damage to the casing occurred.The ifu states ¿improper use of instruments in the handling or implantation of shunt products may result in the cutting, slitting or crushing of components¿.¿.There was proteinaceous debris observed on the interior and exterior of the valve.The ifu also cautions that, ¿care must be taken to ensure that particulate contaminants are not introduced into shunt components during preimplantation testing or handling.Introduction of contaminants could result in improper performance of the shunt system.Particulate matter that enters the shunt system may result in shunt occlusion or may also hold pressure/flow controlling mechanisms open, resulting in overdrainage¿.All valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received reported there were no external/environmental factors identified that may have led to the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATA II VALVE, REGULAR
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key8014302
Report Number2021898-2018-00510
Device Sequence Number1396084
Product Code JXG
UDI-Device Identifier00613994804648
UDI-Public00613994804648
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K042465
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2020
Device Model Number42866
Device Catalogue Number42866
Device Lot NumberE30424
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2018
Supplement Date Received by Manufacturer10/25/2018
11/14/2018
Initial Report FDA Received Date10/29/2018
Supplement Report FDA Received Date10/29/2018
12/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age58 YR
Patient Weight72
-
-