| Catalog Number |
6209000000 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
09/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program. reported events - 1 event was reported for this quarter. product return status - 1 device was received for evaluation. evaluation status - 1 device evaluation is in progress. additional information - 1 device is not labeled for single-use.1 device was not reprocessed and reused.
|
| |
|
Event or Problem Description
|
|
This report summarizes 1 malfunction event in which the device had a component detach.1 event had no patient involvement; no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program. supplemental rationale corrected data 1 event was initially reported.1 previously reported event is included in this follow-up record. product return status 1 device was received.Event confirmation status 1 reported event was confirmed; the cause traced to component failure.Evaluation results 1 device was found to be affected by a missing component.
|
| |
|
Event or Problem Description
|
|
This report summarizes 1 malfunction event in which the device had a component detach.1 event had no patient involvement; no patient impact.
|
| |
|
Search Alerts/Recalls
|