| Catalog Number |
NSLG2C45 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The lot history records were reviewed and the manufacturing criteria were met prior to the release of this lot.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unknown procedure, the tip of the jaw is detached.No patient consequence reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Additional information was requested and the following was obtained: did the detached part fall into the patient? no information was the detached part retrieved from the patient? no information.Did this cause a change in the plan of care for the patient? no information.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # r92z2l.Investigation summary: the device was received with the electrode separated from the ceramic but still attached to the active rod.In addition the jaw was firmly attached and not detached as reported.Upon visual inspection under magnification it was noted that the ceramic was partially missing.The ceramic was damaged by the separation of the electrode from the lower jaw.Testing found a short on the device when the jaws are fully or partially closed, resulting in an alert screen "reposition jaws and reactive".This is advising that the instrument is being activated on low impedance (thin) tissue or metal (such as staples, clips, retractors, or clamps).The "replace instrument" alert screen would be displayed after receiving the "reposition and reactivate" alert screen twice.The separated electrode prevented the instrument from fully cycling open or close.This is due the interference between the blade and electrode.There is insufficient evidence related to what caused the damage on the device.The batch history record was reviewed, and no defects, nc¿s or protocols related to the complaint, were found during the manufacturing process.
|
| |
|
Search Alerts/Recalls
|