| Catalog Number |
5100120952 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr report is part of the malfunction summary reporting program, exemption number e2015055.Reported events - 4 events were reported for this quarter.Product return status - 4 devices were received for evaluation.Evaluation status - 4 device evaluations are in progress.Additional information - 4 devices were not labeled for single-use.4 devices were not reprocessed and reused.
|
| |
|
Event or Problem Description
|
|
This report summarizes 4 malfunction events in which the device reportedly overheated.4 events had no patient involvement; no patient impact.
|
| |
|
Event or Problem Description
|
|
This report summarizes 4 malfunction events in which the device reportedly overheated.4 events had no patient involvement; no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
This mdr report is part of the malfunction summary reporting program, exemption number e2015055.Supplemental rationale corrected data: 4 total events were initially reported.4 previously reported events are included in this follow-up record.Product return status 4 devices were received. evaluation status 2 reported events were confirmed during testing.- 2 devices were found to be affected by a lack of lubrication.2 reported events were not confirmed during testing; however: - 2 devices were found to be affected by internal corrosion.Additional information 4 devices were not labeled for single-use.4 devices were not reprocessed and reused.
|
| |
|