| Catalog Number |
2C8720 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Two single use devices have been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
This report summarizes 2 malfunction events.It was reported that two clearlink y-type blood/solution sets leaked from an unspecified location.This occurred before patient use.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: two single-use devices were received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.All components were found to be correctly placed and per specifications, and no separations between the components was observed.Pressure testing and clear passage testing under water were performed for both devices with no issues noted.The reported condition was not verified or the returned devices.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|