• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO SINGLE TRIGGER ROTARY HANDPIECE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO SINGLE TRIGGER ROTARY HANDPIECE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 6203000000
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 06/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr report is part of the malfunction summary reporting program, exemption number e2015055.Reported events: 2 events were reported for this quarter.Product return status: 2 devices were received for evaluation.Evaluation status: 2 device evaluations are in progress.Additional information: this device is not labeled for single-use.This device was not reprocessed and reused.
 
Event or Problem Description
This report summarizes 2 malfunction events in which the device had a component detach.2 events had no patient involvement; no patient impact.
 
Event or Problem Description
This report summarizes 2 malfunction events in which the device had a component detach.2 events had no patient involvement; no patient impact.
 
Additional Manufacturer Narrative
This mdr report is part of the malfunction summary reporting program, exemption number e2015055.Supplemental rationale corrected data: 2 total events were initially reported.2 previously reported events are included in this follow-up record.  product return status 2 devices were received.  evaluation status confirmed 1 reported event was confirmed during testing.- 1 device was found to be affected by cracked trigger housing.In progress 1 device evaluation is in progress.  additional information 2 devices were not labeled for single-use.2 devices were not reprocessed and reused.
 
Event or Problem Description
This report summarizes 2 malfunction events in which the device had a component detach.2 events had no patient involvement; no patient impact.
 
Additional Manufacturer Narrative
This mdr report is part of the malfunction summary reporting program, exemption number e2015055.Supplemental rationale corrected data: 2 total events were initially reported.2 previously reported events are included in this follow-up record.  product return status 2 devices were received.  event confirmation status 2 reported events were confirmed; the cause traced to component failure.Evaluation results 2 devices were found to be affected by a cracked trigger housing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE TRIGGER ROTARY HANDPIECE
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
MDR Report Key8016106
Report Number0001811755-2018-02232
Device Sequence Number16554091
Product Code HWE
UDI-Device Identifier07613154392339
UDI-Public07613154392339
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized2
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number6203000000
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/02/2018
Supplement Date Received by Manufacturer06/02/2018
06/02/2018
Initial Report FDA Received Date10/29/2018
Supplement Report FDA Received Date01/31/2019
04/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-