CONSOLIDATED MEDICAL EQUIPMENT 1/4"(6.4MM)X6'(1.8M) W/M&F CO; TUBE, ASPIRATING, FLEXIBLE, CONNECTING,
|
Back to Search Results |
|
| Model Number |
0037860 |
| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One (1) (b)(4) was received in original, unopened packaging.The reported catalog and lot numbers were verified.Visual inspection of the device found no obvious abnormalities with regards to the seal of the device however, the poly component of the packaging was stressed at the location in which the white tip of the device meets the packaging.Dye testing was conducted on the device and found that there was no breach of sterility in the seal of the packaging.However, a hole was identified at the location of the stressed poly where ink was able to drip out of the packaging.The hole in the poly component of the packaging compromises the sterility of the device.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution were found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 9 complaints, regarding 10 devices, for this device family and failure mode.During this same time frame 2,466,900 devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).It is standard medical practice to inspect all device prior to use.This issue will continue to be monitored through the complaint system to assure patient safety.
|
| |
|
Event or Problem Description
|
|
During incoming inspection, the distributor rejected this device, (b)(4), ¼"x6' w/m&f co (surgical tubing) lot 201708084 for a tear in the sterile pouch/blister pack.There was no contact with the patient as this was found during incoming inspection.Due to the breach in sterility, this complaint meets the criteria for a reportable event.This report is being raised as a malfunction with potential for injury upon reoccurrence.
|
| |
|
Search Alerts/Recalls
|
|
|