| Catalog Number |
M0035475150 |
| Medical Device Problem Codes |
Material Protrusion/Extrusion (2979); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Thrombosis (2100); No Code Available (3191)
|
| Date of Event |
09/30/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) review confirms that the device met all material, assembly and performance specifications.The subject device was not returned for analysis; therefore, visual, dimension and functional testing could not be performed.In case of this complaint, it is possible that the main coil encountered friction and was stretched during repositioning due to some procedural factors encountered during use due to which vertebral sacrifice had to be performed and medical intervention was required, possibly leading to as reported issues 'nv ¿ main coil stretched', 'nv ¿ coil in catheter friction' and 'nv ¿ patient medical intervention required'; however, it could not be confirmed.As per the additional information provided, no damage was noted to the device prior to use and the device was prepared as per direction for use (dfu) for this product.As assignable cause of undeterminable was assigned to the reported issues, as the device was not returned, and the review and analysis of all available information failed to indicate an assignable cause.
|
| |
|
Event or Problem Description
|
|
It was reported that during coil embolization for a ruptured aneurysm for a patient that was in a poor clinical condition (hunt and hess grade 5), the subject coil was stretched when repositioned.Due to the stretched coil, a vertebral sacrifice had to be performed.No additional information is available at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that during coil embolization for a patient that was in a poor clinical condition ( hunt and hess grade 5), the subject coil was stretched when repositioned.The ruptured aneurysm was located at the basilar artery and the subject coil was protruded from the aneurysm.Due to the stretched coil and thrombus formation, a vertebral sacrifice had to be performed by coil embolization.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) review confirms that the device met all material, assembly and performance specifications.The subject device was not returned for analysis; therefore, visual, dimension and functional testing could not be performed.In case of this complaint it is possible that the main coil encountered friction and was stretched during repositioning due to some procedural factors encountered during use due to which vertebral sacrifice had to be performed and medical intervention was required, possibly leading to the as reported issues.Based on the investigation results and available information an assignable cause of anticipated procedural complication was assigned to the as reported issues 'nv - patient thrombosis (treated vessel) ' and 'nv - patient vessel perforation', since the issues are due to a known physiological effect of the procedure and/or patient condition noted with the direction for use, device labeling and/or risk documentation files.As per the additional information provided by the customer, no damage was noted to the device prior to use and the device was prepared as per direction for use (dfu); therefore, an assignable cause of undetermined was assigned to the as reported issues 'nv main coil stretched', 'nv - patient medical intervention required', and 'nv coil in catheter friction', as the device was not returned, and the review and analysis of all available information failed to indicate as assignable cause.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) review confirms that the device met all material, assembly and performance specifications.The subject device was not returned for analysis; therefore, visual, dimensional and functional testing could not be performed.In case of this complaint, it is possible that the main coil encountered friction and stretched during repositioning due to some procedural factors encountered during use and later protruded due to which a vertebral sacrifice had to be performed and medical intervention was required, possibly leading to as reported issues.Based on the investigation results and available information as assignable cause of anticipated procedural complication was assigned to the as reported issue ' nv- patient thrombosis (treated vessel)', since the issue is due to a known physiological effect of the procedure and/or patient condition noted with the directions for use, device labelling and/or risk documentation files.As per additional information no damage was noted to the device prior to use and the device was prepared as per dfu for this product.An assignable cause of undetermined was assigned to the as reported issues 'nv - main coil stretched', 'nv - patient medical intervention required', 'nv - coil in catheter friction', 'nv - main coil protrusion', as the device was not returned, and the review and analysis of all available information failed to indicate an assignable cause.As reported code nv- patient vessel perforation was updated to nv- main coil protrusion.
|
| |
|
Event or Problem Description
|
|
It was reported that during coil embolization for a patient that was in a poor clinical condition ( hunt and hess grade 5), the subject coil was stretched when repositioned.The ruptured aneurysm was located at the basilar artery and the subject coil was protruded from the aneurysm.Due to the stretched coil and thrombus formation, a vertebral sacrifice had to be performed by coil embolization.
|
| |
|