• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION DREAMTOME RX 44; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION DREAMTOME RX 44; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number M00584060
Medical Device Problem Codes Break (1069); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Visual examination of the returned device revealed that the working length was damaged, the distal pierced hole (tome's extrusion) was found torn, therefore this condition causing the wire anchor to be dislodged from the extrusion, these failure which could have affected the device ability to bow in a clinical setting, confirming the reported complaint.The anchor wire has a section broken, however this section was not returned for analysis and the cutting wire was observed blackened.According to the complaint information, the wire anchor and cutting wire has a section broken and blackened.It is most likely that a peak of voltage could have caused the failures noted.Therefore, a conclusion code of "cause traced to component failure" was assigned to the wire anchor has a section broken, since expected or random component failure without any design or manufacturing issue.However, the device received was found with distal pierce hole torn, this is evidence of that cutting wire anchor slipped causing the catheter to tear and causing the wire anchor to be dislodged from the extrusion.According with the information reviewed in this case there is no evidence of manufacturing issue or mishandle of the device.This type of damage is associated with a known design limitation of the device in which under certain scenarios of usage the catheter can tear, this is an anticipated failure mode within the risk documentation of the product.The most probable cause for the complaint will be assigned as design inadequate for purpose, since the problems traced to design/design features of the device that do not support or interfere with the intended purpose of the device.The dhr review was performed and no anomalies were observed, the review confirmed that the device met all manufacturing specifications and boston scientific manufacturing processes include extensive inspections to ensure that all finished devices meet specifications.
 
Event or Problem Description
It was reported to boston scientific corporation that a dreamtome rx 44 was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2018.According to the complainant, during procedure, when they attempted to perform sphincterotomy, the cutting wire did not bow.Reportedly, there was no visible damage noted on the device.The procedure was completed with a new dreamtome rx 44.There were no user injuries nor patient complications reported as a result of this event.This event has been deemed a reportable event based on the investigation results; cutting wire partially dislodged from the catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMTOME RX 44
Common Device Name
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key8017630
Report Number3005099803-2018-61046
Device Sequence Number11088510
Product Code KNS
UDI-Device Identifier08714729778288
UDI-Public08714729778288
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K013153
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/21/2021
Device Model NumberM00584060
Device Catalogue Number47401
Device Lot Number0022285854
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/02/2018
Initial Report FDA Received Date10/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-