| Catalog Number |
07027885190 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This event occurred in (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer stated that they received erroneous results for one patient sample tested with the elecsys t4 assay on a cobas 8000 e 801 module.The erroneous results were not reported outside of the laboratory.The sample was tested three times on the customer's e 801 analyzer, resulting with t4 values of 24.4 ug/dl, 24.9 ug/dl accompanied by a data flag, and 21.6 ug/dl.The sample was tested at another laboratory on a cobas 6000 e 601 module, resulting with a t4 value of 12.21 ug/dl.No adverse events were alleged to have occurred with the patient.The serial number of the customer's e 801 analyzer is (b)(4).The serial number of the e 601 analyzer used at the other laboratory was asked for, but not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
A sample was provided for investigation.An interfering factor against a component of the reagent was identified.This interference is addressed in the package insert.
|
| |
|