| Catalog Number |
212036 |
| Medical Device Problem Code |
Material Frayed (1262)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If additional information should become available, a supplemental medwatch will be submitted accordingly.(b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
This is report 1 of 2 for the same event.It was reported by the customer in (b)(6) that the suture on the customer's tacit qa anchor and minilok qa anchor broke while in use together.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown but was noted to have occurred in 2018.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device was received and inspected.The complaint can be confirmed.The inserter and suture lengths attached to the needles were received.The inserter was visually inspected to observe any visual anomalies present that may have contributed to the complaint condition, however none were observed.The needles were attached to two suture lengths of approximately the same length with the ends of each suture frayed, indicating that the suture was cut.Per the instructions for use (ifu), overloading the anchor or applying excessive tension to the suture once the anchor is deployed will cause the suture to break therefore is a potential root cause for the reported failure.A non-conformance search was conducted to investigate any defects during production identified that may contribute to the complaint condition.No non-conformance was identified for this part-lot number combination.At this point, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (01)10886705001842(10)l436637(17)220531.
|
| |
|
Event or Problem Description
|
|
Additional event information received on 11/14/2018.Changed legal manufacturer for the 2 products and event occurred during intra-op.
|
| |
|
Event or Problem Description
|
|
Additional event information received on 10/31/2018.It was reported that the event occured at the time of the suture with wire.All fragments were removed from the patient.There is 20 minutes delay (during the capsulodesis with wire pds + implantation of the broach).Patient data is 170 cms and 73 kgs.There were no patient consequences or impact to the user.The procedure was completed when the surgeon changed the operating technique.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device is not being returned, it was discarded by the customer or multiples attempts for product return were made with no response, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.A non-conformance search was performed for this product code 212036 lot l436637 combination and no non-conformances were identified.No further information regarding the technique or instruments used has been provided to determine a root cause for this failure.If any additional information is obtained, this complaint will be re-opened to capture that information.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).Associated medwatch report number: 1221934-2018-55099.
|
| |
|
Search Alerts/Recalls
|