| Model Number |
03.037.112 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2018, the event occurred during a trochanteric femoral nailing system advanced (tfna) procedure.The surgeon had no issue when inserting the tfna and lag screw into the patient.But, when the surgeon removed the aiming arm from the percutaneous handle, it sounded like it cross threaded into the proximal which made it difficult to remove.The surgeon ended up disconnecting the aiming arm, but the connecting screw became temporarily stuck to the nail and then eventually removed.There was a surgical delay of 5 minutes due to attempts in removing the handle.There were no further issues or complications in the procedure.Patient outcome was reported as good.Concomitant devices reported: unknown tfna lag screw (part # unknown, lot # unknown, quantity #1).This report is for one (1) percutaneous radiolucent insertion handle.This is report 1 of 4 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on october 04, 2018, the event occurred during a trochanteric femoral nailing system advanced (tfna) procedure for a femure fracture.The surgeon had no issue when inserting the tfna and lag screw into the patient.But, when the surgeon removed the aiming arm from the percutaneous handle, it sounded like it cross threaded into the proximal which made it difficult to remove.The surgeon ended up disconnecting the aiming arm, but the connecting screw became temporarily stuck to the nail and then eventually removed.There was a surgical delay of 5 minutes due to attempts in removing the handle.There were no further issues or complications in the procedure.Patient outcome was reported as good.Concomitant devices reported: tfna lag screw (part unknown, lot unknown, quantity 1).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a trochanteric femoral nailing system advanced (tfna) procedure on (b)(6) 2018, the surgeon had no issue in inserting the tfna and lag screw into the patient.However, when the surgeon took the aiming arm off in the percutaneous handle, it almost sounded like it was cross threading into the proximal making it difficult to remove.The surgeon ended up removing and getting the aiming arm disconnected.The connecting screw was stuck to nail but eventually removed.The connecting screw was removed by adducting the leg and shaking the handle up and down rapidly until it was removed.There was a surgical delay of 5 minutes trying to find a way to get the handle off.There was no further issues or complications with the surgery.Patient outcome is reported to be fine.Concomitant devices reported: unknown tfna lag screw (part # unknown, lot # unknown, quantity 1).
|
| |
|