| Catalog Number |
01.12.032 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/09/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on 30 october 2018.Lot 081458: (b)(4) items manufactured and released on 10 july 2008.Expiration date: 2013-05-31.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event.Other device involved: liner: cc e cc light flat pe liner ø 28 / e reference 01.26.2844stt.Lot 060220: (b)(4) items manufactured and released on 18 april 2006.Expiration date: 2011-02-28.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Clinical evaluation performed by medica affaris manager: hip revision surgery occurred 9 years after primary cementless total hip arthroplasty.Radiographic images provided show the presence of radiolucent lines around the stem and signs of stress shielding.Aseptic loosening is a possible, literature described adverse event after cementless total hip arthroplasty and causes are often unknown.The reason of this event cannot be determined.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018 we were informed that a revision surgery will be performed due to aseptic femoral loosening, the reason is probably pe wear.
|
| |
|
Additional Manufacturer Narrative
|
|
On 30 october 2018 we were informed that the suspected pe wear is not confirmed.The event description is revision surgery performed, 9 years after primary surgery, due to aseptic femoral loosening.
|
| |
|