• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. METRIX AIR; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. METRIX AIR; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number METRIX AIR
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(manufacturer narrative = t, corrected data = f) (b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi# (b)(4).Note: manufacturer contacted customer on 10/18/2018 in a follow-up call to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention or symptoms since the last call.
 
Event or Problem Description
Consumer reported complaint for low blood glucose test results.Customer is concerned with tests results of 60, 47, 51 and 45 mg/dl.Customer's expected fasting blood glucose test result range is 80 - 120 mg/dl.Customer has only been using the truemetrix air meter for a few weeks.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, back to back blood test was not performed by the customer.Product is stored according to specification in the living area.The test strip lot manufacturer's expiration date is 09/23/2019 and open vial date is one week.The meter memory was reviewed for previous test result history: (b)(6).
 
Additional Manufacturer Narrative
(manufacturer narrative = t, corrected data = f) internal report (b)(4).Returned meter and returned test strips evaluated with no defects found.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip udi# (b)(4).Note: manufacturer contacted customer on (b)(6) 2018 in a follow-up call to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention or symptoms since the last call.
 
Event or Problem Description
Consumer reported complaint for low blood glucose test results.Customer is concerned with tests results of 60, 47, 51 and 45 mg/dl.Customer's expected fasting blood glucose test result range is 80 - 120 mg/dl.Customer has only been using the truemetrix air meter for a few weeks.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, back to back blood test was not performed by the customer.Product is stored according to specification in the living area.The test strip lot manufacturer's expiration date is 09/23/2019 and open vial date is one week.The meter memory was reviewed for previous test result history: result 1:60mg/dl (b)(6) 8:45am fasting, result 2:47mg/dl (b)(6) 9:10am fasting, result 3:51mg/dl (b)(6) 9:10am fasting, result 4:45mg/dl (b)(6) 3:45pm fasting, result 5:80mg/dl(b)(6) 11:09am fasting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METRIX AIR
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key8019422
Report Number1000113657-2018-01114
Device Sequence Number1381903
Product Code NBW
UDI-Device Identifier00041415072787
UDI-Public(01)00041415072787
Combination Product (Y/N)N
Initial Reporter StateFL
PMA/510(K) Number
K150052
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 12/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/23/2019
Device Model NumberMETRIX AIR
Device Catalogue NumberREA4H01-01
Device Lot NumberMV2831
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2018
Was the Report Sent to FDA? No
Event Location Home
Initial Date Received by Manufacturer 10/04/2018
Supplement Date Received by Manufacturer10/04/2018
Initial Report FDA Received Date10/30/2018
Supplement Report FDA Received Date12/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SECOND THERAPY; SECOND THERAPY
-
-