MAUDE Adverse Event Report: COOK IRELAND LTD ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE; FCG KIT, NEEDLE, BIOPSY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK IRELAND LTD ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE; FCG KIT, NEEDLE, BIOPSY Back to Search Results
Model Number G53585
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/09/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
510 (k) number: k142688.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.Investigation is still pending.A follow up mdr will be submitted to include the investigation conclusions.
 
Event or Problem Description
As reported to customer relations (cr), "the doctor reports difficulty in maneuvering the handle and when puncturing and retracting the needle it got stuck, in both needles the same happens.No patient complication.The doctor decided to use another needle of the boston brand.Both needles were used only and they did not achieve tissue since they were tubed.The needle is revised and we observe that the needle of both devices has a burr.
 
Event or Problem Description
As reported to customer relations (cr), the doctor reported difficulty in manouvering the handle and when puncturing and retracting the needle it gets stuck, in both needles the same happens.No patient complication.The doctor decided to use another needle of the boston brand.Both needles were used only and they did not achieve tissue since they were tubbed.The needle is revised and we observe that the needle of both devices has a burr.
 
Additional Manufacturer Narrative
510 (k) number; k142688 cook (b)(4) ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031 information pertaining to section g.1 as follows: importer site contact and address: (b)(6) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195 importer site establishment registration number: 3005580113 investigation is still pending.A follow up mdr will be submitted to include the investigation conclusions.
 
Additional Manufacturer Narrative
510 (k) number; k142688.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4).Cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.The two devices related to this occurrence underwent a laboratory evaluation.The needle tips on both devices were found to be bent.Prior to distribution, all echo-hd-19-c devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for echo-hd-19-c of lot number c1427326 did not reveal any discrepancies that could have contributed to this complaint issue.The review of relevant manufacturing records, confirms the failure mode has not previously occurred with the current lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1427326.The notes section of the instructions for use, which accompanies this device instructs the user to inspect the device prior to use : "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use".There is no evidence to suggest that the customer did not follow the instructions for use.A definitive root cause for the customer complaint could not be determined as circumstances of use cannot be replicated in the laboratory.A possible cause could be attributed to advancement into a hard lesion.Complaint is confirmed based on the customer's testimony.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event or Problem Description
As reported to customer relations (cr), the doctor reported difficulty in manoeuvering the handle and when puncturing and retracting the needle it gets stuck, in both needles the same happens.No patient complication.The doctor decided to use another needle of the boston brand.Both needles were used only and they did not achieve tissue since they were tubed.The needle is revised and we observe that the needle of both devices has a burr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHOTIP PROCORE HIGH DEFINITION ULTRASOUND BIOPSY NEEDLE
Common Device Name
FCG KIT, NEEDLE, BIOPSY
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
MDR Report Key8019855
Report Number3001845648-2018-00504
Device Sequence Number1395960
Product Code FCG
UDI-Device Identifier00827002535855
UDI-Public(01)00827002535855(17)201130(10)C1427326
Combination Product (Y/N)N
Initial Reporter CountryMX
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2020
Device Model NumberG53585
Device Catalogue NumberECHO-HD-19-C
Device Lot NumberC1427326
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2018
Was the Report Sent to FDA? No
Distributor Facility Aware Date10/30/2018
Event Location Hospital
Initial Date Received by Manufacturer 10/03/2018
Supplement Date Received by Manufacturer10/03/2018
10/03/2018
Initial Report FDA Received Date10/30/2018
Supplement Report FDA Received Date11/27/2018
12/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
-
-