• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US ANTEROMEDIAL FEMORAL AIMER 7.5MM -NS; ORTHOPAEDIC IMPLANT AIMING/GUIDING BLOCK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US ANTEROMEDIAL FEMORAL AIMER 7.5MM -NS; ORTHOPAEDIC IMPLANT AIMING/GUIDING BLOCK Back to Search Results
Catalog Number 254687
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If additional information should become available, a supplemental medwatch will be submitted accordingly.(b)(4).Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported by the sales rep via phone that while checking his inventory, the sales rep's anteromedial femoral aimer 7.5mm has imperfections around the exit point.The sales rep stated the wire does not pass through easily and can create metal shavings.The issue was found unrelated to a case, and therefore there was no patient involvement and no delays.The device is being returned for evaluation.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If additional information should become available, a supplemental medwatch will be submitted accordingly.Investigation summary: the complaint device was received and evaluated.Visual examination of the distal end of the device revealed some scratches at the edge of the inner pass at the distal end.The complaint of metal shavings can be confirmed.The possible root cause for the observed damage could be the device used with a pin which was slightly bent.Other than above mentioned possibility, we cannot discern any definite root cause at this point of time.Review of the device history record indicated that this batch of product was processed without incident; therefore there is no evidence of manufacturing anomalies on the records reviewed.At this time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTEROMEDIAL FEMORAL AIMER 7.5MM -NS
Common Device Name
ORTHOPAEDIC IMPLANT AIMING/GUIDING BLOCK
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key8019933
Report Number1221934-2018-55112
Device Sequence Number1756180
Product Code LXH
UDI-Device Identifier10886705013685
UDI-Public10886705013685
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254687
Device Lot Number1805001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2018
Initial Date Received by Manufacturer 10/14/2018
Supplement Date Received by Manufacturer12/03/2018
12/03/2018
Initial Report FDA Received Date10/30/2018
Supplement Report FDA Received Date12/05/2018
12/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-