• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY CHARGER; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY CHARGER; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 1600US
Medical Device Problem Code Device Misassembled During Manufacturing /Shipping (2912)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.Product event summary: the battery charger was returned for evaluation.A review of the manufacturing documentation confirmed that the associated device met all requirements for release.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.Failure analysis of the returned battery charger revealed that the unit passed visual inspection and functional testing.The reported battery charger "flashing red" could not be confirmed during testing.However, supplemental testing revealed that the battery charger was assembled with incorrect inductors.The incorrect inductors might over charge a battery, causing the battery to flash red on the battery charger due to an eight (8) hour charge time-out.As a result, the most likely root cause of the reported event can be attributed to improper assembly.An internal investigation was initiated to investigate battery chargers assembled with incorrect inductors.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported, by product analysis,that the incorrect inductor was used during the assembly process.The battery charger was exchange.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY CHARGER
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key8020295
Report Number3007042319-2018-05235
Device Sequence Number1380780
Product Code DSQ
UDI-Device Identifier00888707000130
UDI-Public00888707000130
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/14/2018
Device Model Number1600US
Device Catalogue Number1600US
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/18/2018
Initial Report FDA Received Date10/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-