Catalog Number 221832460
Medical Device Problem Codes Naturally Worn (2988); Scratched Material (3020)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Complaint not specific to one case.Surgeons complain that pinnacle poly trials are worn and used, scratches on them and screws worn out.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).No device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review, was not possible because the required lot code(s) was not provided.The investigation could not draw any conclusions regarding the reported event.Based on the inability to determine a root cause, the need for corrective action was not indicated.Product received 11 dec 2018.Examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot = null.Device history batch = null.Device history review = null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIN TRL LNR +4 NEUT 600DX32ID
Common Device Name
HIP INSTRUMENTS : ACETABULAR TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8020691
Report Number1818910-2018-73888
Device Sequence Number16489071
Product Code LXH
UDI-Device Identifier10603295101079
UDI-Public10603295101079
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number221832460
Device Lot NumberCY0201
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2018
Initial Date Received by Manufacturer 10/15/2018
Supplement Date Received by Manufacturer12/11/2018
Initial Report FDA Received Date10/30/2018
Supplement Report FDA Received Date12/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-