Catalog Number ST16LN50NGF151S
Medical Device Problem Code Noise, Audible (3273)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Event or Problem Description
Information was received indicating that a patient with a smiths medical portex® bivona® adult tts¿ tracheostomy tube makes a whistling noise with breathing.A backup tracheostomy (trach) was reported to be on hand.There were no reported adverse patient effects.
 
Additional Manufacturer Narrative
One tracheostomy returned for evaluation.The device was reviewed and it was found to meet all of the dimensional requirements.The product was visually inspected and found to have no abnormalities, as well as leak tested.It has passed all functional testing.The reported customer complaint was not confirmed.While no definitive root cause to the reported issue could be determined, this investigation revealed no intrinsic evidence to suggest a root cause related to manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX BIVONA FLEXTEND¿ TTS¿ TRACHEOSTOMY TUBES
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD; INC.
6000 nathan lane n
minneapolis, MN 55442
MDR Report Key8020717
Report Number3012307300-2018-01938
Device Sequence Number13751791
Product Code JOH
UDI-Device Identifier15019517090529
UDI-Public15019517090529
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K923878
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date08/16/2023
Device Catalogue NumberST16LN50NGF151S
Device Lot NumberDS018253
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/02/2018
Supplement Date Received by Manufacturer11/15/2018
Initial Report FDA Received Date10/30/2018
Supplement Report FDA Received Date12/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-