| Model Number |
10798696 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.
|
| |
|
Event or Problem Description
|
|
The customer reported that the tubing set leaked at the posterior end of the filter while infusing a tpn via the patient's peripheral iv.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer¿s complaint of filter leaking was confirmed.Visual inspection showed no anomalies.A leak was observed between the tubing and the distal end of the filter.The tubing measurement was within iso-specification.The root cause of the customer¿s report of leaking from the filter was due to insufficient solvent at the engagement between the filter outlet and tubing.
|
| |
|
Event or Problem Description
|
|
The customer reported that while in nicu the tubing set leaked at the posterior end of the filter while infusing a tpn via the patient's peripheral iv.There was no patient harm.
|
| |
|