| Catalog Number |
133MV-14-T |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Device labeling: adverse reactions may require premature explantation, which may affect desired flap size.Patients should be advised that the tissue expanders may deflate, and require replacement surgery.Deflation occurs when saline leaks through a damaged injection site or a damaged tissue expander envelope.
|
| |
|
Event or Problem Description
|
|
Healthcare provider reported right side "ruptured" and capsular contracture baker grade ii.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: opening, red particles, fluids clear inside the device.Leak test was performed and found opening on anterior.A microscopic analysis was performed which identified a striated edge opening, consist with use of a surgical tool.Based on the device analysis the final assessment is: a striated edge opening on anterior side due to surgical damage.
|
| |
|
Event or Problem Description
|
|
Healthcare provider reported right side "ruptured" and capsular contracture baker grade ii.Device has been explanted.
|
| |
|