• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 133MV TISSUE EXPANDER WITH SUTURE TAB; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 133MV TISSUE EXPANDER WITH SUTURE TAB; EXPANDER, SKIN, INFLATABLE Back to Search Results
Catalog Number 133MV-14-T
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Device labeling: adverse reactions may require premature explantation, which may affect desired flap size.Patients should be advised that the tissue expanders may deflate, and require replacement surgery.Deflation occurs when saline leaks through a damaged injection site or a damaged tissue expander envelope.
 
Event or Problem Description
Healthcare provider reported right side "ruptured" and capsular contracture baker grade ii.Device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified: opening, red particles, fluids clear inside the device.Leak test was performed and found opening on anterior.A microscopic analysis was performed which identified a striated edge opening, consist with use of a surgical tool.Based on the device analysis the final assessment is: a striated edge opening on anterior side due to surgical damage.
 
Event or Problem Description
Healthcare provider reported right side "ruptured" and capsular contracture baker grade ii.Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 133MV TISSUE EXPANDER WITH SUTURE TAB
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key8021178
Report Number9617229-2018-08344
Device Sequence Number19177854
Product Code LCJ
UDI-Device Identifier10888628004436
UDI-Public10888628004436
Combination Product (Y/N)N
PMA/510(K) Number
K102806
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/13/2022
Device Catalogue Number133MV-14-T
Device Lot Number3175839
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2018
Initial Date Received by Manufacturer 10/05/2018
Supplement Date Received by Manufacturer11/14/2018
Initial Report FDA Received Date10/30/2018
Supplement Report FDA Received Date12/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-