| Catalog Number |
150419 |
| Medical Device Problem Code |
Restricted Flow rate (1248)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the valve on the leg bag did not allow urine to flow out.The patient had to manipulate the tubing in order pass the urine through.No medical intervention was required.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use was found adequate and states the following: "1.Separate notches within circles.Pull straps through holes and around leg.2.Position bag on leg with flutter valve at top.3.Attach catheter or extension tubing to top inlet.When wearing bag below knee, attach bard extension tubing (catalog no.150615 or 4a4194).4.To empty dispoz-a-bag, push green lever on flip-flo valve out and down.Important: be sure to reclose flip-flo valve after emptying bag.5.After use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable, local, state and federal laws and regulations.".
|
| |
|
Event or Problem Description
|
|
It was reported that the valve on the leg bag did not allow urine to flow out.The patient had to manipulate the tubing in order pass the urine through.No medical intervention was required.
|
| |
|