| Catalog Number |
5530-G-513 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
10/02/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that: "right knee revision, one component (poly).Mechanical complication of right knee.
|
| |
|
Event or Problem Description
|
|
It was reported that: "right knee revision, one component (poly).Mechanical complication of right knee.Update (b)(6) 2018 from med review: x-rays shows posterior subluxation of femur and varus instability.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding revision due to joint instability involving a triathlon insert was reported.It was noted on an image provided of the explanted component that a section of the insert fractured.The event was confirmed through the x-rays and photographs provided.Method & results: device evaluation and results: no product was returned for evaluation however one photograph was provided.The photograph shows a recently explanted insert.Damage is visible on one of the two compartments on the posterior edge of the bearing surface.Medical records received and evaluation: a review of the provided medical records were rejected for medical review by a clinical consultant the following comment was provided: x-rays shows posterior subluxation of femur and varus instability; need operative reports, past/clinical medical history.Device history review: could not be performed as lot code information was not provided.Complaint history review: could not be performed as lot code information was not provided.Conclusion: the damage on the component was confirmed on the photograph of the explanted device and instability was confirmed through review of the x-rays provided however, the exact cause of the event could not be determined because insufficient information was provided.Further information such as product evaluation,operative reports, past/clinical medical history are needed to complete the investigation for determining a root cause.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Search Alerts/Recalls
|