MAUDE Adverse Event Report: DEPUY MITEK LLC US EXPRESSEW III W/O HOOK; SUTURE/NEEDLE PASSER, REUSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US EXPRESSEW III W/O HOOK; SUTURE/NEEDLE PASSER, REUSABLE Back to Search Results
Catalog Number 214140
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/10/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement: the complaint device was received and inspected.Visual observations revealed the jaw pin was protruding out of its space contributing to the jaw failure and misalignment of the jaws.The damage to the jaw pin would result in needle deployment issues, confirming this complaint.Clamping excessive tissue between the jaws combined with repetitive twisting action exerts excess side load on the jaw pin resulting in the jaw pin shearing off from its space.This failure can be attributed to user technique.A batch record review has been conducted and our results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy synthes mitek complaints system revealed one other dissimilar complaint for this lot of devices that were released to distribution.Based on the overall complaint rate and customer impact, at this point in time, no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.This report is being filed from the (b)(4) complaint management system as required under mitek¿s corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event or Problem Description
The sales representative reported that during a rotator cuff repair that the customer's expressew iii without hook was not firing the needle correctly.The sales representative reported that the end of the device was slightly bent.The surgeon completed the procedure with another like device with no patient consequences.There was a ten (10) minute delay reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESSEW III W/O HOOK
Common Device Name
SUTURE/NEEDLE PASSER, REUSABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key8023647
Report Number1221934-2018-51282
Device Sequence Number13798348
Product Code LXH
UDI-Device Identifier10886982153821
UDI-Public10886982153821
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number214140
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2016
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/17/2016
Initial Report FDA Received Date10/31/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-