Catalog Number 5552-L-350
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Codes Fall (1848); Injury (2348)
Date of Event 10/02/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the product history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
Sales rep reported, "patient fell and landed in a kneeling position.Left patella broke.Fractured patella repaired and tritanium asymmetric patella revised to a poly asymmetric patella.".
 
Additional Manufacturer Narrative
An event regarding crack/fracture involving a tritanium patella was reported.The event was confirmed through x-ray.Product evaluation and results: not performed as no product was returned for evaluation, no photograph was provided for review.A review of the provided medical records by a clinical consultant stated the following comment: confirmation of the patella fracture and event description.Indicated all devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other similar events for the lot referenced.A review of the provided medical records by a clinical consultant stated the following comment: confirmation of the patella fracture and event description.The exact cause of the event could not be determined because insufficient information was provided.Further information such as device return, pre and post operative x-rays, operative reports as well as clinical/past patient history and follow up notes are required to complete the investigation for determining a root cause.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Sales rep reported, "patient fell and landed in a kneeling position.Left patella broke.Fractured patella repaired and tritanium asymmetric patella revised to a poly asymmetric patella.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRITANIUM PATELLA-ASYMMETRIC
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8023753
Report Number0002249697-2018-03559
Device Sequence Number11100876
Product Code MBH
UDI-Device Identifier07613327013559
UDI-Public07613327013559
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2023
Device Catalogue Number5552-L-350
Device Lot NumberE00Y
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/02/2018
Supplement Date Received by Manufacturer11/27/2018
Initial Report FDA Received Date10/31/2018
Supplement Report FDA Received Date12/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age55 YR
Patient Weight124
-
-