| Catalog Number |
VR496 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial suture breakage malfunction events: 226.Initial suture breakage devices returned: 246.17 use error caused or contributed to event 192 no device failure.21 manufacturing deficiency.16 unable to confirm complaint.Initial needle breakage malfunction events: 102.Initial needle breakage devices returned: 161.20 use error caused or contributed to event.129 no device failure.3 manufacturing deficiency.9 unable to confirm complaint.Initial needle pull off malfunction events: 147.Initial needle pull off devices returned: 254.15 use error caused or contributed to event.159 no device failure.48 manufacturing deficiency.32 unable to confirm complaint.Follow-up suture breakage malfunction events: 16.Follow-up suture breakage devices returned: 41.1 use error caused or contributed to event.39 no device failure.0 manufacturing deficiency.1 unable to confirm complaint.Follow-up needle breakage malfunction events: 9.Follow-up needle breakage devices returned: 7.1 use error caused or contributed to event 1 no device failure.3 manufacturing deficiency.2 unable to confirm complaint.Follow-up needle pull off malfunction events: 17.Follow-up needle pull off devices returned: 43.5 use error caused or contributed to event.32 no device failure.4 manufacturing deficiency.2 unable to confirm complaint.All devices labeled for single use; 0 devices reprocessed and re-used 0 remedial action.Complaint information will be included in complaint trending that is reviewed on a regular basis to determine if further action is necessary.
|
| |
|
Event or Problem Description
|
|
This report summarizes 517 malfunction events, including 242 suture breakage, 111 needle breakage and 164 detachment of device component that occurred intraoperatively.
|
| |
|
Search Alerts/Recalls
|