Batch review performed on 31 october 2018.Lot 1720581: (b)(4) items manufactured and released on 05 january 2018.Expiration date: 2022-12-10.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Visual inspection performed the 30th of october 2018 by r&d product manager: only the implant (pedicle screw) was returned.The implant looks intact without signs of defect.The event seem related to "uncomfort" with the design/ergonomics of the instrument; however, no failure/damage/"breackage" of the same can be assessed.Alternative designs of compression devices are available and more under development.Other device involved: pedicle screw compression pliers, reference 03.51.10.0041, lot.Unknown.
|