MEDTRONIC IRELAND VALIANT CAPTIVIA - FF; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number VAMF4646C150TU
Medical Device Problem Codes Difficult or Delayed Positioning (1157); Inadequacy of Device Shape and/or Size (1583); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Intimal Dissection (1333)
Date of Event 10/12/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A valiant stent graft system was implanted in a patient for the endovascular treatment of a 58mm diameter thoracic aortic aneurysm.It was reported that during the index procedure after the deployment of the device the delivery system got stuck in the external iliac.It was stated that the physician performed a cut down at the area where the delivery system was stuck.The delivery system was removed and the artery was repaired with a synthetic graft.The event was resolved.Per the physician, the cause of the event was due to patient's anatomy; patients artery was too small for the size of the delivery system.No additional clinical sequelae were reported and the patient is fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALIANT CAPTIVIA - FF
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key8024012
Report Number2953200-2018-01520
Device Sequence Number1093322
Product Code MIH
UDI-Device Identifier00643169346918
UDI-Public00643169346918
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P100040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/12/2020
Device Model NumberVAMF4646C150TU
Device Catalogue NumberVAMF4646C150TU
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/12/2018
Initial Report FDA Received Date10/31/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient Weight64
-
-