| Model Number |
M002000195140 |
| Medical Device Problem Code |
Product Quality Problem (1506)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
10/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
To date, the serial number of the concerned patch is not available, only the lot number was confirmed.
|
| |
|
Event or Problem Description
|
|
During a carotid endarterectomy, a first patch was implanted but looked "more yellow" and it felt "fuzzier and thicker" than previous used patches (see manufacturer report 1640201-2018-00021 for this first involved device).The surgeon opened a second box (sterile pack remained closed) with a different lot number.This second patch looked the same and is the subject of the present report.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The returned device was visually inspected by our qa supervisor.No anomaly was found, the patch complies with our product specifications.(b)(4) the conducted investigation indicates that the product was not defective.
|
| |
|