BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 381033
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device expiration date: unknown.Device manufacture date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that a bd insyte¿ autoguard¿ bc shielded iv catheter had a burr on the catheter tip.There was no report of exposure, serious injury, or medical intervention.
 
Additional Manufacturer Narrative
Investigation summary: the lot number associated with this investigation was unknown therefore, dhr review could not be performed.Received a 20ga iag bc unit with no packaging material.All components were intact.Visual/microscopic evaluation: white-clear material was observed near the tip of the catheter tubing of the unit received.Conclusion(s): the white-clear material observed on the catheter tubing was confirmed to be non-foreign (plug shavings) and to have resulted from manufacturing during the assembly process.
 
Event or Problem Description
It was reported that a bd insyte¿ autoguard¿ bc shielded iv catheter had a burr on the catheter tip.There was no report of exposure, serious injury, or medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE¿ AUTOGUARD¿ BC SHIELDED IV CATHETER
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
MDR Report Key8025237
Report Number1710034-2018-00770
Device Sequence Number16441613
Product Code FOZ
UDI-Device Identifier00382903810338
UDI-Public382903810338
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K110443
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number381033
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2018
Initial Date Received by Manufacturer 10/11/2018
Supplement Date Received by Manufacturer10/11/2018
Initial Report FDA Received Date10/31/2018
Supplement Report FDA Received Date11/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-