• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. VISIONAIRE JII CUTTING BLOCK KIT- RIGHT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. VISIONAIRE JII CUTTING BLOCK KIT- RIGHT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number V0100083
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/10/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that surgeon downsized femur from planned."not" injury reported.
 
Additional Manufacturer Narrative
The associated journey ii visionaire cutting block kit was not returned for evaluation.Therefore a product evaluation could not be performed.However, case details have been provided.Therefore, an evaluation was conducted based on the information received.The review of the manufacturing records for the listed batch did not reveal any deviation from the standard manufacturing processes.Our investigation including an engineering evaluation by our visionaire team noted that after evaluation, , no root cause could be found for the complaint.All parts of the design were within visionaire standards.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIONAIRE JII CUTTING BLOCK KIT- RIGHT
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8025319
Report Number1020279-2018-02321
Device Sequence Number13726426
Product Code JWH
UDI-Device Identifier00885556492673
UDI-Public00885556492673
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K082358
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date01/19/2019
Device Catalogue NumberV0100083
Device Lot Number00119189V1
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/10/2018
Supplement Date Received by Manufacturer10/10/2018
Initial Report FDA Received Date10/31/2018
Supplement Report FDA Received Date05/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-