| Catalog Number |
319.04 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that on an (b)(6) 2018, the depth gauge was found broken by medical device reprocessing department (mdrd).The silver hook appears to be unthreaded from the black piece.There was no patient involvement.This report is for one (1) depth gauge for 2.7mm & small screws.This is report 1 of 1 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Customer quality conducted an investigation of the returned device.Device evaluation: broken.Visual inspection observed that the needle portion of the depth gauge was broken where the needle articulates with the slider of the depth gauge.There is slight surface wear present on the outside of the guide sleeve.The received condition was determined to agree with the complaint condition.During the document/specification review the following drawings, reflecting the current revision, were reviewed.Drawings reflecting the manufactured date of the device were not reviewed as the lot number for the device is unknown.Depth gauge for 2.7mm to 4.0mm screws: slider assembly depth gauge: no definitive cause of the issue could be determined.Dimensional inspection: dimensional inspection of needle component could not be conducted due to post manufacturing damage.No review of the device history record(s) or material and hardness review could be conducted due to the lot number being unknown.The exact cause of the complaint condition cannot be determined as the handling and use of the device are unknown.However, the condition of the depth gauge is consistent with the result of a bending force applied to the needle portion of the depth gauge that is beyond the yield limit of the material.Investigation conclusion: the complaint condition is confirmed as the needle was observed to be broken from the slider of the depth gauge.No design or manufacturing issues were identified.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The device was received, the investigation is in progess, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|