Model Number ZXR00
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pt info: unknown/ not provided.Date of event: unknown, not provided; best estimate is between (b)(6) and (b)(6) 2018.(b)(4).Attempts have been made to obtain additional information; however, to date a response has not been received.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a zxr00 29.5 diopter intraocular lens (iol) was implanted in the patient's operative eye on (b)(6) 2018.It was later explanted on (b)(6) 2018, due to a calculation issue.There was no other surgical intervention required.The replacement lens was a zxr00 24.0, a smaller diopter lens.No additional information was provided.
 
Additional Manufacturer Narrative
Device evaluation: product testing could not be performed because the product was not returned.The customer's reported event could not be confirmed.Manufacturing records review: during the manufacturing record review of the production order for this serial number, no product deficiency was identified.The documentation shows that the production order was manufactured according to specifications.Historical data analysis: a search of complaints revealed no other complaints have been received for this production order number.Labeling review: the labeling review was completed and revealed that the directions for use(dfu) adequately provide instructions, precautions, along with warnings for the proper use and handling of the product.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON AND JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key8026155
Report Number9614546-2018-01028
Device Sequence Number16481931
Product Code HQL
UDI-Device Identifier05050474591783
UDI-Public(01)05050474591783(17)211102
Combination Product (Y/N)Y
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/04/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/02/2021
Device Model NumberZXR00
Device Catalogue NumberZXR00U0295
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 10/04/2018
Supplement Date Received by Manufacturer11/20/2018
10/25/2020
Initial Report FDA Received Date10/31/2018
Supplement Report FDA Received Date11/30/2018
11/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1MTEC30 CARTRIDGE, LOT NUMBER UNKNOWN
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-