It was reported that the caller stated that when the device was interrogated the device was elective replacement indicator (eri) with settings of vvi 65.After uploading the programmer data to the patient management database application, the atrial ohms were displayed as 67 ohms.The atrial lead was off due to the device being at eri and possibly due to an erroneous number during the data transfer.The caller was advised to test the lead during a generator change to ensure the atrial lead was functioning normally.The application remains in use.There was no patient involvement.
|