Related to 1218996-2018-00022, user facility called (b)(4) product support team to report discrepant patient results.A pediatric patient was ran on the leadcare ii blood lead testing system and gave a result of 3.3g/dl.When this same patient was sent to a reference lab for a venous confirmatory test the result was 8.0g/dl.The customer's leadcare ii blood lead test kit was returned to magellan for evaluation.During investigative testing, no abnormal results were recorded.It should be noted during the investigation, visual inspection of the returned leadcare ii analyzer confirmed corrosion on the sensor pins.Corrosion of the sensor pins occurs when there is overexposure of the analyzer pins to the assay's acidic treatment reagent.This occurs when used test sensors are not removed immediately after testing (per instrument instructions) or when too much sample is added to the test sensor.Periodic testing with blood lead qc controls is used to monitor analyzer performance.(b)(4) provided the customer with a new instrument, and additional training.No reported harm or injuries occurred as a result of this issue.(b)(4).
|