| Model Number |
ZXR00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Halo (2227); No Code Available (3191)
|
| Date of Event |
06/11/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Device evaluation: the product testing could not be performed as the product was not returned.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to specifications.A search revealed that no similar complaints were received for this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a tecnis symfony multifocal intraocular lens (model zxr00, +13.5 diopter) was explanted from patient¿s operative eye because of the patient experiencing intolerable glare and halos.The replacement lens was a monofocal, model pcb00.Incision enlargement was performed.There was no sutures and no vitrectomy performed.Reportedly visual acuity pre-op was 20/30 and visual acuity post-op: 20/20.Intraocular lens was exchanged with resolutions in the symptoms.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
| |
|
Search Alerts/Recalls
|