Model Number ZXR00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Halo (2227); No Code Available (3191)
Date of Event 06/11/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Device evaluation: the product testing could not be performed as the product was not returned.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to specifications.A search revealed that no similar complaints were received for this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision has been submitted.
 
Event or Problem Description
It was reported that a tecnis symfony multifocal intraocular lens (model zxr00, +13.5 diopter) was explanted from patient¿s operative eye because of the patient experiencing intolerable glare and halos.The replacement lens was a monofocal, model pcb00.Incision enlargement was performed.There was no sutures and no vitrectomy performed.Reportedly visual acuity pre-op was 20/30 and visual acuity post-op: 20/20.Intraocular lens was exchanged with resolutions in the symptoms.No further information provided.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key8027108
Report Number9614546-2018-01032
Device Sequence Number11159412
Product Code POE
UDI-Device Identifier05050474591714
UDI-Public(01)05050474591714(17)220609
Combination Product (Y/N)Y
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/04/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/09/2022
Device Model NumberZXR00
Device Catalogue NumberZXR00U0135
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 10/02/2018
Supplement Date Received by Manufacturer10/25/2020
Initial Report FDA Received Date10/31/2018
Supplement Report FDA Received Date11/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
Date Report Sent to FDA01/01/2005
-
-