| Model Number |
694775 |
| Medical Device Problem Codes |
Fracture (1260); High impedance (1291)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/18/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead had fractured.The lead had high undefined impedance on the superior vena cava (svc) coil.The lead was removed and a new lead was implanted.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the partial lead was returned in segments and analyzed.The analysis indicated that the interior superior vena cava defibrillation cable developed a fracture due to flexing while in vivo.If information is provided in the future, a supplemental report will be issued.
|
| |
|