WARSAW ORTHOPEDICS CD HORIZON SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Catalog Number 6430530
Medical Device Problem Code Failure to Align (2522)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Pre-operative diagnosis: spinal canal stenosis procedure: posterior fusion after oblique lumbar interbody fusion (olif) levels implanted: l3-5 it was reported that intra-op, when trying to perform final tightening from one side of l3,the set screw idled.Product came in contact with the patient.Patient complications were reported as unknown.
 
Additional Manufacturer Narrative
Product analysis: visual and macroscopic examination did not reveal thread crest/flank damage but some damage to the hex of the screw.It appears the hex was damaged from being over-torqued.Functional evaluation with a sample mas head found the set screw was able to be fully engaged.The above observations are consistent with torsional overload.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key8027286
Report Number1030489-2018-01438
Device Sequence Number203684
Product Code NKB
UDI-Device Identifier00643169545281
UDI-Public00643169545281
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K143375
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number6430530
Device Lot Number0601779W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2019
Initial Date Received by Manufacturer 10/15/2018
Supplement Date Received by Manufacturer02/26/2019
Initial Report FDA Received Date11/01/2018
Supplement Report FDA Received Date03/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-