JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS; MONOFOCAL IOLS Back to Search Results
Model Number ZA9003
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Vitrectomy (2643); No Code Available (3191)
Date of Event 10/03/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Implant and explant dates: if implanted or explanted, give date: not applicable, as the lens was inserted and removed (b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the doctor noticed a little bit of blood in the patient's operative eye, and then when a za9003 +16.5 diopter intraocular lens was being implanted, the doctor encountered more blood.Therefore, the lens was removed, an anterior vitrectomy was performed and sutures were used.The patient went home aphakic and will be referred to a retina specialist.No additional information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation: yes, returned to manufacturer on: 10/26/2018, device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection using magnification was performed.The lens returned with residues of viscoelastic on the optic body.Both lens haptics were observed detached.Only one detached haptic piece was returned.The condition observed is consistent with a lens that has been handling during surgical process.The reported issue could not be verified.No product deficiency was identified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no additional investigation request were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key8027776
Report Number2648035-2018-01417
Device Sequence Number1262995
Product Code HQL
UDI-Device Identifier05050474528949
UDI-Public(01)05050474528949(17)230510
Combination Product (Y/N)Y
Initial Reporter StateMD
PMA/510(K) Number
P990080
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/04/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/10/2023
Device Model NumberZA9003
Device Catalogue NumberZA90030165
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2018
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 10/04/2018
Supplement Date Received by Manufacturer11/21/2018
10/25/2020
Initial Report FDA Received Date11/01/2018
Supplement Report FDA Received Date11/25/2018
11/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EMERALDC30 CARTRIDGE, LOT UNKNOWN.; EMERALDC30 CARTRIDGE, LOT UNKNOWN.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
Date Report Sent to FDA01/01/2005
-
-