| Model Number |
D284TRK |
| Medical Device Problem Codes |
Failure to Charge (1085); Delayed Charge Time (2586)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device was reported as included in the field action.Based on the information received and without the return of the product, it cannot be confirmed that this device performed as described in the field action.It is included in the field action in the abundance of caution.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) alerted for a charge circuit timeout which occurs when charge times exceed a certain threshold.It was noted that the patient had already received a heart transplant and the device was no longer providing any therapy.The device remains in the patient.No patient complications have been reported as a result of this event.
|
| |
|