| Catalog Number |
LINC9254 |
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
09/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Since the product was not returned for analysis, no product failure analysis can be conducted and no determination of possible contributing factors could be made.The device history record (dhr) was review and no discrepancies related to the reported issue were uncovered.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) year old male patient underwent a shoulder arthroplasty performed with a sterlimed shaver "shaver blade sterling cuda large hub magenta" and metal shavings were observed during the procedure.It was unknown what intervention was performed.The device was exchanged and the procedure was continued.The issue with the metal shavings observed during the procedure was assessed as reportable.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received on november 1, 2018 confirming that there was no patient injury.Manufacturer's reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to 3500a initial report.Manufacturer state code¿ field reflecting as ca and it should be mn.Manufacturer contact state code was left blank and it should be mn.Manufacturer site state code was left blank and it should be mn.These fields, have been processed accordingly.Manufacturer's reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: it was reported that a 45 year old male patient underwent a shoulder arthroplasty performed with a sterlimed shaver "shaver blade sterling cuda large hub magenta" and metal shavings were observed during the procedure.It was unknown what intervention was performed.The device was exchanged and the procedure was continued.There was no patient injury reported.The device was returned loose within a resealable plastic bag.None of the original packaging or label was returned.There were no obvious visual damage or deformities observed.There are minor biological contaminants still present on the distal area of the shaver- indicative of handling within or outside of the operative field.The cutting edges appear sharp and clean; there are no observed pitting, nicks, or metal fatigue that could explain the presence of shavings.There are no metal shavings observed to be stuck to any of the still present biological matter.The inside surface area of the outer piece was examined and there is no observed damage or scouring marks that could result in metal shavings.The outer and inner sub-assemblies were taken apart.The shafts of both pieces appear straight.The two parts were reassembled together and the drop/spin test was conducted.The inner piece spun freely and there was no observed metal-on-metal rubbing.The assemblies were wiped down with light alcohol, using a lint-free cloth to collect any residue that could product metal shavings.No shavings were observed or collected.While there are indications of handling, no conclusion as to the cause of the reported issue is determined.The reported issue is not verified.The device history record (dhr) was review and no discrepancies related to the reported issue were uncovered.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The biosense webster inc.Failure analysis lab received the device for evaluation on 11/1/2018.Therefore, populated device available for evaluation?, is device returned to manufacturer? and date device returned to manufacturer.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.Manufacturer's ref.No: (b)(4).
|
| |
|