Advanced bionics considers the investigation into this reportable event as closed.Prior to explant surgery, external equipment was exchanged and programming adjustments were made, however, the issue did not resolve.The external visual inspection revealed that the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented some of the electrical tests from being performed.The device passed the electrical and mechanical tests performed.The residual gas analysis could not be performed due to the ceramic case cracking while attempting to pull vacuum.The reported complaint of intermittent malfunction could not be verified during this analysis, which was limited in some respects due to the electrode array being cut prior to receipt, as well as damage induced during the residual gas analysis.This device passed all of the electrical tests performed.This older device configuration is no longer manufactured.This is the final report.
|