BARD ACCESS SYSTEMS POWERPICC PROVENA CATHETER WITH SHERLOCK 3CG TIP POSITIONING SYSTEM (TPS) STYLET; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number S1385108D
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a bent stylet is confirmed but the exact cause is unknown.One photo sample of a 3cg stylet was returned for investigation.The stylet was observed to have a curve at the proximal section of the polyimide tubing and a kink at the end of the curve.Based on the photo sample provided, possible contributing factors include advancement and retraction against resistance.Since a bend and a kink were observed on the stylet, the complaint is confirmed but the exact cause remains unknown.A lot history review (lhr) of recs0928 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported that after successful placement of a provena catheter, the tps wire showed an unusual bend in the distal portion after removal.10/09/2018 - returned wire is kinked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC PROVENA CATHETER WITH SHERLOCK 3CG TIP POSITIONING SYSTEM (TPS) STYLET
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8028681
Report Number3006260740-2018-03067
Device Sequence Number1421398
Product Code LJS
UDI-Device Identifier00801741112720
UDI-Public(01)00801741112720
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K053501
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 11/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberS1385108D
Device Catalogue NumberS1385108D
Device Lot NumberRECS0928
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2018
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/09/2018
Initial Report FDA Received Date11/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-