| Catalog Number |
3549316 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun, but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.The device history record (dhr) was reviewed, and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a mentor cpx 4 breast tissue expander with suture tabs 750cc had a hole when the box was opened.The hole was discovered pre-operation so there was no patient contact.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that a mentor cpx 4 breast tissue expander with suture tabs 750cc had a hole when the box was opened.The hole was discovered pre-operation so there was no patient contact.The device was returned to mentor without fluid inside.No foreign material was observed within the device or on the shell surface.The mentor product analysis lab discovered a rent measuring approximately 2.4 cm on the posterior aspect.Microscopic examination of the edges of the rent revealed parallel striations.No other anomalies were discovered.The complaint was confirmed since a rupture was found in the device.Microscopic examination revealed parallel striations.This type of striations is more conclusive to sharp instrument damage rather than shell failure due to wear.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet.(b)(4).
|
| |
|
Search Alerts/Recalls
|