PENUMBRA, INC. RUBY COIL; HCG, KRD Back to Search Results
Catalog Number RBY2C2457
Medical Device Problem Codes Physical Resistance/Sticking (4012); Premature Separation (4045)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report numbers: 3005168196-2018-02179; 3005168196-2018-02180; 3005168196-2018-02182; 3005168196-2018-02183.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the splenic artery using a pod12, pod packing coils (podjs) and ruby coils.It was noted that the patient¿s anatomy was tortuous.During the procedure, while attempting to deploy a pod12 into the target vessel using a lantern delivery microcatheter (lantern), the physician determined that the pod12 was oversized and therefore, it was removed.The pod12 pusher assembly then became kinked when the hospital technologist submerged it into saline to flush it.The physician then decided to remove the lantern.Next, the physician implanted a pod10 in the target vessel using a non-penumbra microcatheter.While advancing a podj through the microcatheter, the physician experienced resistance and the pusher assembly of the podj became kinked on the back end of the microcatheter; therefore, it was removed.The physician then implanted another podj in the target vessel with no issues.While advancing a ruby coil through the microcatheter, the physician experienced resistance and subsequently, the ruby coil unintentionally detached, with some of the ruby coil in the patient's body.Therefore, the physician used manual aspiration to pull the detached coil inside the microcatheter and removed the microcatheter and coil together.The ruby coil was then removed and the microcatheter reinserted into the patient¿s body.After that, the physician successfully placed another ruby coil in the target vessel.While advancing a podj through the microcatheter, resistance was encountered and the podj pusher assembly became kinked; therefore, it was removed.The physician then placed another podj in the target vessel with no issues.While advancing another podj through the microcatheter, resistance was also encountered and the podj pusher assembly became kinked; therefore, it was removed.The procedure was completed using a new podj.It should be noted that a third microcatheter was used, but it is unknown at what point in the procedure.There was no alleged deficiency against the lantern; the physician preferred to use another microcatheter.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Results: the pet lock was wrinkled but intact on the proximal end of the pusher assembly.An embolization coil was not attached to the pusher assembly but the proximal constraint remained loaded in the distal detachment tip (ddt).A stretch resistant component was not visible in the proximal constraint sphere.The pusher assembly was undamaged.Conclusions: evaluation of the returned pod12 confirmed a kinked device.The complaint reported that the pod12 was kinked when submerging the device into saline after retracting it outside the patient.If the device is forcefully mishandled outside of the patient, damage such as a kink may occur.During functional testing, the pod12 was able to advance and retract through its introducer sheath and a demo lantern without issue.Evaluation of the podj confirmed a kinked device.The complaint reported that resistance was experienced when advancing the podj through a non-penumbra microcatheter.If the device is forcefully advanced against resistance, damage such as a kink may occur.Based on the reported complaint, some of the kinks are likely incidental to the complaint and may have occurred post procedure.During functional testing, the device was able to be advanced through its introducer sheath and a demonstration lantern with resistance due to the pusher assembly kinks.The non-penumbra microcatheter was not returned for evaluation; therefore, the root cause of the initial resistance was unable to be determined.Evaluation of the ruby coil confirmed the embolization coil was missing from the pusher assembly.The proximal constraint sphere was present within the ddt.There was not a stretch resistant component visible in the constraint sphere, therefore, the root cause of the coil becoming detached could not be determined.The embolization coil was not returned for evaluation; therefore, the device was unable to be functionally tested.Evaluation of the podj was unable to confirm the reported kinks.Further evaluation revealed a detached embolization coil and the ddt was sheared off the distal end of the pusher assembly.If the ddt becomes sheared off the pusher assembly, it will likely cause the embolization coil to become detached.Based on the reported complaint, the sheared ddt and detached embolization coil were likely incidental.Evaluation of the podj confirmed a kinked device.If the device is forcefully advanced against resistance, the pusher assembly may become kinked.The non-penumbra microcatheter was not returned for evaluation; therefore, the root cause of the resistance was unable to be determined.Further evaluation revealed additional kinks and a detached embolization coil.Based on the reported complaint and return condition, these damages were likely incidental to the complaint.No other devices identified in the complaint were returned for evaluation.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The investigation findings do not lead to a clear conclusion about the cause of coil becoming detached.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 1.3005168196-2018-02179, 2.3005168196-2018-02180, 3.3005168196-2018-02182, and 4.3005168196-2018-02183.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8028958
Report Number3005168196-2018-02181
Device Sequence Number1585458
Product Code HCG
UDI-Device Identifier00814548019291
UDI-Public00814548019291
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K173614
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/04/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/11/2026
Device Catalogue NumberRBY2C2457
Device Lot NumberF84016
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 10/04/2018
Supplement Date Received by Manufacturer06/06/2019
Initial Report FDA Received Date11/01/2018
Supplement Report FDA Received Date06/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age63 YR
Date Report Sent to FDA01/01/2005
-
-