| Model Number |
D334DRG |
| Medical Device Problem Code |
Delayed Charge Time (2586)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/11/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the implantable cardioverter defibrillator (icd) exhibited excessive charge time and went straight to end of service (eos).The device is expected to be replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.However, performance data collected from the device was received and analyzed.Analysis of the device memory indicated an alert for excessive charge time eos (end of service).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was further reported that the device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was returned and analyzed.Analysis of the device found high internal battery resistance.Device analysis confirmed that the device did not charge efficiently with the original device battery.The device charged nominally when the battery was replaced with an external supply.No hybrid anomalies were found.The results were consistent with the device battery causing the excessive charge time.The reported excessive charge times likely resulted from a spurious increased battery resistance induced by lithium-severing, a known failure mode for 35 joule batteries.If information is provided in the future, a supplemental report will be issued.
|
| |
|